Standard Operating Procedure (SOP)
1. Objective
To standardize the chemical fogging and defogging process in controlled environments to ensure effective disinfection and compliance with microbial control standards.
2. Scope
This SOP applies to the following areas:
- Microbial Limit Testing Room
3. Responsibilities
- Microbiology Team: Perform fogging/defogging, maintain records.
4. Materials & Equipment
- Fogging Machine: Wet Fogger
- Disinfectant: 2% W/V Care P-AS Solution
- PPE: Gloves, goggles, mask, shoe covers, protective gown
5. Procedure
5.1 Pre-Fogging Preparations
- Notify relevant departments before starting.
- Prepare disinfectant solution as per table:
Total Volume (mL) | Care P-AS (mL) | Purified Water (mL) |
---|---|---|
1000 | 20 | 980 |
2000 | 40 | 1960 |
- Fill the fogger tank and ensure proper assembly.
- Display warning signs at entry points.
5.2 Fogging Execution
- Set fogging parameters per room:
Room | Disinfectant (mL) | Time (min) |
---|---|---|
MLT room | 30 | 2 |
Change room | 60 | 4 |
- Adjust fogger nozzle and flow rate (0–9 scale).
- Start fogging using the digital timer (DG 60).
- After completion, leave the room closed for 60 minutes.
5.3 Defogging Process
- Activate AHU (heater OFF) for 30 minutes.
- Open all doors (except Change Room-I).
- Ensure no personnel remain in the area.
5.4 Post-Fogging Activities
- Clean the fogger with purified water.
- Document details in Annexure-01.
6. Safety Precautions
- Mandatory PPE: Gloves, goggles, mask, shoe covers.
- Strictly prohibit entry during fogging/defogging.
- Emergency: Rinse eyes/skin with water if exposed; seek medical help.
7. Abbreviations
Term | Full Form |
---|---|
MB | Microbiology |
AHU | Air Handling Unit |
HEPA | High-Efficiency Particulate Air |
8. References
- Entry/Exit Procedure for Microbial Areas
- Disinfectant Preparation & Cleaning
9. Annexures
- Annexure-01: Fogging & Defogging Record
Fogging & Defogging Record
Field | Details |
---|---|
Date of Procedure: | |
Performed By: | |
Room/Area: | |
Fogging Machine Serial No.: | |
Disinfectant Volume (mL): | |
Fogging Time (Min): | |
Fogger Nozzle Settings (0–9): | |
Flow Rate Setting (0–9): | |
Timer Used (Yes/No): | |
AHU Activated (Heater OFF for 30 Min): | |
All Doors Opened (except Change Room-I): | |
No Personnel in Area: | |
Fogger Cleaned with Purified Water: | <